• Users Online: 1062
  • Print this page
  • Email this page


 
 Table of Contents  
CASE REPORT
Year : 2023  |  Volume : 7  |  Issue : 1  |  Page : 15-16

A rare complication of intra-aortic balloon pump


1 Department of Adult Intensive Care, Prince Sultan Cardiac Center, Riyadh, Saudi Arabia
2 Department of Adult Cardiac Surgery, Prince Sultan Cardiac Center, Riyadh, Saudi Arabia
3 Adult Cardiac Surgery, Prince Sultan Cardiac Center, Riyadh, Saudi Arabia; Department of Cardiothoracic Surgery, Tanta University, Tanta, Egypt

Date of Submission27-Nov-2022
Date of Acceptance24-Dec-2022
Date of Web Publication28-Mar-2023

Correspondence Address:
Amr A Arafat
Department of Adult Cardiac Surgery, Prince Sultan Cardiac Center, Building 6, Makkah Al Mukarramah Branch Road, As Sulimaniyah, Riyadh 12233

Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/sccj.sccj_28_22

Rights and Permissions
  Abstract 


Intra-aortic balloon pump (IABP) is commonly used for circulatory support in patients with cardiogenic shock. IABP is associated with several complications; however, IABP rupture was rarely reported. We present a female patient with acute myocardial infarction and ventricular septal rupture. The patient underwent transcatheter device closure of the ventricular septum and IABP insertion. After 4 days of insertion, we encountered a failure to cycle alarm. The IABP was removed with difficulty, and a leak coming from the shaft near the bifurcation was found. This case highlighted the importance of early detection of IABP rupture to avoid entrapment and further complications.

Keywords: Intra-aortic balloon, percutaneous mechanical support devices, rupture


How to cite this article:
Aboughanima MA, ElSayed A, Arafat AA. A rare complication of intra-aortic balloon pump. Saudi Crit Care J 2023;7:15-6

How to cite this URL:
Aboughanima MA, ElSayed A, Arafat AA. A rare complication of intra-aortic balloon pump. Saudi Crit Care J [serial online] 2023 [cited 2023 Jun 7];7:15-6. Available from: https://www.sccj-sa.org/text.asp?2023/7/1/15/372682




  Introduction Top


Percutaneous circulatory support plays a vital role in managing cardiogenic shock after acute myocardial infarction.[1] Intra-aortic balloon pump (IABP) improves left ventricular unloading and augments diastolic coronary blood flow.[2] Despite the modest effect of IABP compared with other circulatory support devices, IABP is simple to use and has a good safety profile. On the other hand, IABP is associated with several complications, including limb ischemia, rupture, and entrapment.[3] We reported a case of IABP rupture that was managed with early IABP removal.


  Case Report Top


Female patients aged 78 years had acute myocardial infarction complicated by ventricular septal rupture. The patient had urgent transcatheter device closure of the ventricular septal defect and intra-aortic balloon insertion (Maquet, San Jose, CA, US). After 4 days of IABP insertion, we encountered a failure to cycle alarm. There was no leak in the connection, and the mark of the helium container indicated an adequate level. Despite that, we changed the console twice.

Balloon rupture was a possibility, but we could not detect any blood in the helium compartment of the tube. The treating team decided to take the balloon out. However, there was difficulty in removing the balloon. Repeated aspiration and negative pressure were used to deflate the balloon, and a sheath was inserted to facilitate removal without injuring the artery. The balloon was taken out and no blood clots were detected on the balloon.

We put the balloon underwater and checked for any leaks, but there was no leak. However, we noticed that on injecting air, the balloon does not inflate very well, and applying pressure on the balloon leads to a loss of its volume [[Figure 1] and Video 1]. We put the whole balloon shaft underwater to find a leak coming from the shaft near the bifurcation. The patient had a smooth recovery with no complications related to the IABP insertion site.
Figure 1: The intra-aortic balloon pump after removal showing leakage

Click here to view



  Discussion Top


Despite the safety profile of IABP compared to other mechanical-assisted devices, it is associated with major complications that can be life-threatening.[3] The reported complications with IABP include limb ischemia, aortic dissection, bleeding, and rarely rupture and entrapment.[4],[5]

In our patient, IABP rupture was diagnosed due to a malfunctioning balloon with failure to inflation. In such cases, the treating team should check for any leak from connections or empty hilum containers, which were negative in our case. Although there was no blood inside the lumen, balloon rupture was highly suspected. Therefore, the team decided to remove the IABP to avoid potential complications. IABP rupture could lead to entrapment, and forceful removal might damage the iliac or femoral arteries.[6] The balloon came out from our patient with difficulty but without vascular or limb complications. Few cases were published about managing entrapped IABP; however, our case shows that early recognition and prompt removal using sheaths are vital to avoid complications during removal. Delayed diagnosis could lead to balloon entrapment and increase the possibility of surgical removal of the balloon. Unusual sites of ruptures, such as the shaft, should be suspected when no blood is found inside the balloon.


  Conclusions Top


Early diagnosis and removal of the ruptured intra-aortic balloon are essential to avoid vascular complications and the need for surgical removal.

Declaration of patient consent

The authors certify that they have obtained all appropriate patient consent forms. In the form the patient(s) has/have given his/her/their consent for his/her/their images and other clinical information to be reported in the journal. The patients understand that their names and initials will not be published and due efforts will be made to conceal their identity, but anonymity cannot be guaranteed.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Wernly B, Seelmaier C, Leistner D, Stähli BE, Pretsch I, Lichtenauer M, et al. Mechanical circulatory support with impella versus intra-aortic balloon pump or medical treatment in cardiogenic shock – A critical appraisal of current data. Clin Res Cardiol 2019;108:1249-57.  Back to cited text no. 1
    
2.
Ahmad Y, Sen S, Shun-Shin MJ, Ouyang J, Finegold JA, Al-Lamee RK, et al. Intra-aortic balloon pump therapy for acute myocardial infarction: A meta-analysis. JAMA Intern Med 2015;175:931-9.  Back to cited text no. 2
    
3.
Parissis H, Soo A, Al-Alao B. Intra aortic balloon pump: Literature review of risk factors related to complications of the intraaortic balloon pump. J Cardiothorac Surg 2011;6:147.  Back to cited text no. 3
    
4.
Lewis PA, Mullany DV, Townsend S, Johnson J, Wood L, Courtney M, et al. Trends in intra-aortic balloon counterpulsation: Comparison of a 669 record Australian dataset with the multinational benchmark counterpulsation outcomes registry. Anaesth Intensive Care 2007;35:13-9.  Back to cited text no. 4
    
5.
Siriwardena M, Pilbrow A, Frampton C, MacDonald SM, Wilkins GT, Richards AM. Complications of intra-aortic balloon pump use: Does the final position of the IABP tip matter? Anaesth Intensive Care 2015;43:66-73.  Back to cited text no. 5
    
6.
Sabzi F, Zokaei AH, Moloudi AR, Dabiri S. Intra-Aortic balloon pump entrapment in a nonatherosclerotic common iliac artery. Open J Cardiovasc Surg 2011;4:5-7.  Back to cited text no. 6
    


    Figures

  [Figure 1]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Case Report
Discussion
Conclusions
References
Article Figures

 Article Access Statistics
    Viewed698    
    Printed36    
    Emailed0    
    PDF Downloaded63    
    Comments [Add]    

Recommend this journal


[TAG2]
[TAG3]
[TAG4]